Open original FDA record ↗Recall record
- FDA identifier
- Z-2394-2016
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Aug 11, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The possibility exists for contaminants, in the form of biomass, to develop in the cooling system of Artis systems resulting in damage to the pump system.
Source facts
- Action
- Siemens has implemented a Customer Safety advisory Notice to all affected customer via Update Instruction AX007/15/S. Additional, a corrective action will be implemented via Update Instructions AX006/15/S and AX003/16/S.
- Initiated
- 2016-07-13
- Root cause
- Under Investigation by firm
- Terminated
- 2017-11-08
- Distribution
- Nationwide Distribution to: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WV,WY, Puerto Rico and Guam
- Recall number
- Z-2394-2016
- Recall status
- Terminated
- CDRH record ID
- 148281
- FDA record created
- 2016-08-11
- Recall event number
- 74716