Open original FDA record ↗Recall record
- FDA identifier
- Z-2394-2023
- Company in FDA record
- Stryker Sustainability Solutions
- FDA classification date
- Aug 11, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
39-inche-wide inflatable air transfer mattress packages contain 34-Inch-wide mattresses, so there may be a decreased surface area available to distribute the patient's weight across, which may increase the effort exerted by staff for a lateral patient transfer, and there may be increased friction and shearing forces when the patient is moved resulting in skin tear injury.
Source facts
- Action
- On 6/21/23, recall notices were distributed to customers who were asked to do the following: 1) Return affected devices. 2) Complete and return the Business Reply form to SSSPFA@stryker.com 3) If any of the affected product have been forwarded to additional facilities (including but not limited to distribution warehouses), contact these facilities and communicate this recall to them and instruct them to complete and return the Business Reply Form. Adverse reactions or quality problems can be reported to the Complaint Hotline: +1-888-763-8803 or http://www.stryker.com/productexperience
- Initiated
- 2023-06-21
- Root cause
- Packaging process control
- Distribution
- US: WI, ID, CA, FL, HI, MD, RI, AZ, MA, GA, IL
- Recall number
- Z-2394-2023
- Recall status
- Open, Classified
- CDRH record ID
- 201972
- FDA classified
- 2023-08-11
- Recall event number
- 92655