- FDA identifier
- Z-2395-2008
- Company in FDA record
- Synthes U S A
- FDA classification date
- Sep 21, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The product is mislabeled, as the validation for the previous sterilization parameters cannot be replicated. The currently validated sterilization parameter is a minium of 24 minutes prevacuum.
Source facts
- Action
- All Synthes Trauma customers and sales consultant s were notified of the correction via return receipt letters sent via UPS. This notification communication contains product literature documenting the validated sterilization parameters. Instructions detail disposal of the previously received literature and replacement with the provided literature. A toll free number with a dedicated extension has been established for questions. The firm has requested that all recipients acknowledge understanding of instructions by returning a Verification Section via fax or email. If you have questions, call 800-620-7025 ext 5450, 610-719-5450 or contact your Synthes Sales Consultant.
- Initiated
- 2008-07-23
- Root cause
- Error in labeling
- Terminated
- 2009-05-05
- Distribution
- The product was shipped to medical facilities and sales consultants nationwide. The product was also shipped to Canada.
- Recall number
- Z-2395-2008
- Recall status
- Terminated
- CDRH record ID
- 72569
- FDA classified
- 2008-09-21
- Recall event number
- 49039