Open original FDA record ↗Recall record
- FDA identifier
- Z-2396-2012
- Company in FDA record
- US Endoscopy Group Inc
- FDA classification date
- Sep 17, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
US Endoscopy received five complaints which noted difficulty in deploying the cytology brush through the catheter.
Source facts
- Action
- US Endoscopy sent an Addendum and Correction to the Instructions for Use letter dated December 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to rinse the brush head with sterile water immediately prior to use (reference the attached Addendum. Customers were also instructed to retract and deploy the brush to ensure that is is functioning properly prior to use. Customers were asked to review the instruction for use for this product along with the Addendum to ensure proper usage and performance. For questions customers should call 1-410-639-4494 or their local Endoscopic Product Specialist. For questions regarding this recall call 440-639-4494, ext 378.
- Initiated
- 2011-12-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-03-22
- Distribution
- Worldwide Distribution - USA including CA, FL, GA, IA, IL, MA, ME, MN, MO, NC, NH, OH, OK, PA, TN, TX, VA, WA & WV and the countries of Australia, Canada, Columbia, Finland, Germany, India, Israel, Italy, Netherlands, New Zealand & United Kingdom.
- Recall number
- Z-2396-2012
- Recall status
- Terminated
- CDRH record ID
- 112365
- FDA classified
- 2012-09-17
- Recall event number
- 63047