An Ad-Tech Clinical Specialist, attended a case on April 15, 2019. During the case the surgeon encountered an issue
after implanting four Ad-Tech depth electrodes. Upon initial device testing subsequent to implantation it was identified
that two of the devices that were labeled as 8 contact depth electrodes (SD08R-SP05X-000) were in fact 6 contact
depth electrodes (SD06R-SP05X-000) . There was no impact to the patient due to this incident.
Source facts
Action
Via a MEDICAL DEVICVE RECALL letter dated, April 17, 2019, Ad-Tech advised their consignees of the mix-up, requested that they check their stock, returned the MEDICAL DEVICE RECALL RETURN RESPONSE, and return the subject devices. Distributors were requested to conduct a sub-recall to their customers.
Initiated
2019-04-16
Root cause
Labeling Change Control
Terminated
2022-05-27
Distribution
Worldwide - Nationwide distribution in the state of Illinois
and countries of Russia and United Kingdom
Recall number
Z-2396-2019
Recall status
Terminated
CDRH record ID
172558
FDA record created
2019-08-26
Recall event number
82719
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.