Z-2397-2008: Centros(TM) 15F x 28cm Chronic Hemodialysis Curved Catheter Set (With Cuff 23cm From Tip), Catalog No./RE

Recall record Z-2397-2008 for Angiodynamics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2397-2008
Company in FDA record
Angiodynamics, Inc.
FDA classification date
Sep 24, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The cuff may be inadequately attached to the catheter, resulting in possible catheter movement or leakage at the insertion site.

Source facts

Action
AngioDynamics sent URGENT MEDICAL DEVICE letters and reply forms to the consignees via Certified Mail on/about 7/10/08. The recall is being conducted to the medical user level. AngioDynamics is requesting the return of all Centros Hemodialysis Catheters in inventory. The following steps should be taken by the consignees: 1. Identify and segregate the recalled product that is in your possession. 2. Complete the enclosed Centros Hemodialysis Catheter Recall Reply Form and fax it to the attention of the Centros Recall Coordinator at 518-798-1360. It is important that even if you do not have any product remaining in your possession that you fill out the attached form noting zero quantity to be returned and fax the form to AngioDynamics. 3. Ship the recalled product(s) to AngioDynamics. Consignees will be credited for the returned products. Credits will only be issued for returned product that is within the scope of the recall. Reference Return Authorization Number ___ on the outside of the shipping box and include a copy of the Centros Recall Reply Form with the return shipment. --- The letters also provide actions to be taken for product(s) in use: ANGIODYNAMICS recommends that physicians continue to follow facility protocols for inserting dialysis catheters and monitoring dialysis catheter patients. It is recommended that physicians consider evaluation of patients with an indwelling Centros catheter to assure the catheter is properly secured. As stated in the product's instructions for use, it is recommended that the catheter be secured/sutured for the entire duration of implantation. If the sutures have been cut, consider re-suturing to maintain proper catheter position. Do not rely on ingrowth of the cuff to maintain proper catheter position. This failure does not affect the function or integrity of the catheter itself and there have no reports of patient injury due to this problem. For questions, concerns, or more detailed instructions on how to
Initiated
2008-07-07
Root cause
Process design
Terminated
2012-08-02
Distribution
Nationwide Distribution
Recall number
Z-2397-2008
Recall status
Terminated
CDRH record ID
72578
FDA classified
2008-09-24
Recall event number
49042

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗