Z-2397-2019: Phadia" 1000 Model # 12-3800-01 Product Usage: The instrument is fully integrated and automated instrume

Recall record Z-2397-2019 for Phadia Ab. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2397-2019
Company in FDA record
Phadia Ab
FDA record creation date
Aug 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

code 7-102 Liquid Sensor Error

Source facts

Action
The following actions are to be taken: 1. Use of the Phadia" 1000 can continue as detailed in the users manual with the following changes: If the error 7-102 occurs, ensure that the error is handled as stated below: A. Select Retry if the error occurs outside of Assay run. This selection has no effect on assay performance in this situation. B. Select Retry if the error occurred within Assay run in the previous six minutes. This selection has no effect on assay performance in this situation. Note: If you are unsure of the six minutes related to the error occurrence, select Stop. C. Select Stop if the error occurred within Assay run after six minutes as indicated in the system software. Note: Performing a Retry or Stop after six minutes presents a risk that assay performance and results are affected. Selecting Stop will stop ImmunoCAP dispensing and all processing tests will be flagged as erroneous. Corrective actions can then be performed. 2. Whether you select Retry or Stop, contact Technical Support after attending to error 7-102 according to the instructions above to provide notification to our technical department and to receive additional guidance. Additionally; Ensure that the sound alarm is audible and that the visible alarm is visible within the laboratory. Ensure that the criteria according to the user manual for incoming water (rinse solution) is met. Further, it is recommended to ensure that Monthly Maintenance with 1% Sodium Hypochlorite (bleach) solution is performed according to the user manual. Ensure Rinse and Wash Bottle Float/Stems are correctly installed in their respective bottles and are refilled in a timely manner when the instrument is operated. Ensure Stop Solution bottles are full at the start of Assay run, otherwise switch bottle connector to second bottle loaded prior to the first becoming empty. Note: Stop Solution bottles contain 1200 doses. If you have any questions, please
Initiated
2019-07-09
Root cause
Under Investigation by firm
Terminated
2022-07-12
Distribution
US Nationwide Distribution in the states of CA CO FL GA HI IL IN KS KY MA MD MI MO NC NJ NV NY OH OK OR PA TN TX UT VA WA
Recall number
Z-2397-2019
Recall status
Terminated
CDRH record ID
175079
FDA record created
2019-08-26
Recall event number
83402

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗