Open original FDA record ↗Recall record
- FDA identifier
- Z-2399-2020
- Company in FDA record
- Leica Microsystems, Inc.
- FDA record creation date
- Jun 17, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Possible unintended interruption of surgical procedures due to equipment shutdown, related to changes in voltage.
Source facts
- Action
- The firm sent letters to consignees on May 20, 2020 advising them that they can continue to use the system pending this field update. Based on our data analysis, the probability of occurrence for this issue (excessive voltage fluctuation and a subsequent microscope shut off) is remote. However, please ensure that a functional check is performed prior to each surgery as described in section 16.1, directions for the operator of the instrument of the user manual.
- Initiated
- 2020-05-20
- Root cause
- Other
- Terminated
- 2021-05-13
- Distribution
- Domestic distribution: AL,AR,AZ, CA, CT, DC, DE, FL, IL, KY, LA, MI, MN, MO, NC, NY, OH. OR. TN, TX, UT, VA, WA, WI ,WV. International distribution: Japan
- Recall number
- Z-2399-2020
- Recall status
- Terminated
- CDRH record ID
- 181626
- FDA record created
- 2020-06-17
- Recall event number
- 85722