Open original FDA record ↗Recall record
- FDA identifier
- Z-2399-2026
- Company in FDA record
- DT MedTech, LLC
- FDA classification date
- Jun 10, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Through customer complaints, the ankle replacement system was found to contain incorrect devices: Left implant packaging and labeling bears a right implant device.
Source facts
- Action
- On March 24, 2026, DT MedTech issued a Urgent Medical Device Recall Notification to affected consignees. DT MedTech requested that the firm take the following actions: 1. Return of all available stock of Hintermann Series Talar Implants (Part/REF number 302112, Lot number AACAA) 2. Please provide confirmation that all available stock of the part detailed below has been returned to Vilex LLC at 4516 S 700 E Suite 100A Salt Lake City, Utah 84107 USA. 3. For any questions or concerns, contact Vilex Quality at quality@vilex.com or 800-521-5002.
- Initiated
- 2026-03-24
- Root cause
- Labeling Change Control
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, AR, AZ, ID and the countries of PER, KOR.
- Recall number
- Z-2399-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220366
- FDA classified
- 2026-06-10
- Recall event number
- 98987