Z-2400-2020: Mueller Knife Handle, Hexagonal - Product Usage: Intended to hold surgical blades.

Recall record Z-2400-2020 for Carefusion 2200 Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2400-2020
Company in FDA record
Carefusion 2200 Inc
FDA record creation date
Jun 18, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Becton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets (Figure 1) contain a machining burr, potentially causing the collet to not fully seat in the knife handle (Figure 2), inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade.

Source facts

Action
Becton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets contain a machining burr, potentially causing the collet to not fully seat in the knife handle, inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade. BD is asking consignees to: Please check all inventory locations within your institution for the affected Catalog No/Lot number combination listed in the recall notice. Discard all affected product (both collet and handle assembly) following your institutions process for discarding. If discarding is not immediately possible at your facility, then product should be quarantined until disposal. 2. This recall should be carried out to the user level. If you have further distributed any of the affected product, please contact that location immediately and advise them of the recall, forward these instructions and have them complete the instructions within this recall notification. 3. Share this recall notification with all users of the product within your institution to ensure they are also aware of this recall. 4. Complete the attached Customer Response Form and return it to the contact noted on the form regardless of whether you have any affected product or not so that BD may acknowledge your receipt of this notification. Indicate on the response form the quantity of affected product identified at your facility and confirm that this inventory has been discarded. Credit will be issued to customers for all discarded product. 5. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program.
Initiated
2020-05-18
Root cause
Under Investigation by firm
Terminated
2021-04-25
Distribution
US Nationwide distribution including in the states of CA, FL, VA and KY.
Recall number
Z-2400-2020
Recall status
Terminated
CDRH record ID
181673
FDA record created
2020-06-18
Recall event number
85747

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗