Z-2401-2020: Alinity C, Processing Module. Chemistry analyzer for in-vitro diagnostics.

Recall record Z-2401-2020 for Abbott Gmbh & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2401-2020
Company in FDA record
Abbott Gmbh & Co. KG
FDA record creation date
Jun 18, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential to generate incorrect patient results for the following assays: Alkaline Phosphatase, Amylase, Creatine Kinase, and Gamma Glutamyl Transferase.

Source facts

Action
Abbott sent a notification dated May 12, 2020 titled, "Product Correction Urgent Immediate Action Required," to customers. This Product Correction letter is to inform customers of an issue with the Alinity c system Optics Assembly (part number 3501627801). The firm has identified the potential for a nonlinear optics response due to an issue with the Alinity c optics. The issue is with the diffraction grating within the optics assembly and can result in a lower than expected absorbance value. The root cause of the issue is unknown and currently under investigation. The issue can be identified using an Optics diagnostic procedure. Instruments that pass this Optics diagnostic procedure do not exhibit and will not develop the nonlinear optics response. Abbott has tested multiple representative assays with different calibration types and wavelengths to understand the impact of the nonlinear optics response on the firm's clinical chemistry menu. The studies compared sample results obtained between an Alinity c with a confirmed normal optics to the results obtained from an Alinity c with confirmed defective optics. The studies show that four assays are impacted by the nonlinear optics response: Alkaline Phosphatase, Amylase, Creatine Kinase, and GammaGlutamyl Transferase. The impact on the test result is a false decrease in results. An Appendix is attached to the notification regarding the assay testing information. There is a potential to generate incorrect patient results for the following assays on the Alinity c system: Alkaline Phosphatase, Amylase, Creatine Kinase, and GammaGlutamyl Tansferase. Abbott recommends discontinuing testing on the impacted assays (Alkaline Phosphatase, Amylase, Creatine Kinase, and GammaGlutamyl Transferase) until an assessment of the optics can be completed. A firm representative will contact the customer to schedule the assessment to determine the potential impact to the optics assembly. If customers have forwarded t
Initiated
2020-05-12
Root cause
Component design/selection
Terminated
2023-01-31
Distribution
U.S. States: AR, CA, CO, FL, GA, KS, KY, LA, MD, MI, MN, MO, MS, NC, NJ, NY, OK, SC, SD, TN, TX, UT, VA, WI, PR Worldwide distribution: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BERMUDA, BRAZIL, CANADA, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAMAICA, JORDAN, KENYA, MALAYSAI, MEXICO, NEW CALEDONIA, NORWAY, OMAN, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDIA ARABIA, SERBIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, URUGUAY, and VIETNAM.
Recall number
Z-2401-2020
Recall status
Terminated
CDRH record ID
181543
FDA record created
2020-06-18
Recall event number
85686

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗