Dilator tips may break in the package and in patients during surgical procedures.
Source facts
Action
A customer notification letter was issued to customers on May 19, 2023. Customers were requested to identify affected product; cease and quarantine any product manufactured prior to December 31, 2019.
Acknowledgement of receipt should be made to Olympus per the instructions in the recall letter.
Initiated
2023-05-19
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - United States Nationwide and the countries of South Korea, Canada, Australia, Taiwan, Hong Kong, Japan, and Germany.
Recall number
Z-2401-2023
Recall status
Open, Classified
CDRH record ID
201158
FDA classified
2023-08-14
Recall event number
92495
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.