Reinforcement spring may detach from the cannula body resulting in reduced venous blood return flow.
Source facts
Action
All domestic consignees were notified by a Product Recall letter on 07/15/2008 via FedEx. They were instructed to check their inventory for any affected product and return all unused product from their stock to Edwards. Customers are to contact Customer Service Organization to obtain a Returned Goods Authorization number and replacement product. Once inventory has been verified, the attached form needs to be completed and faxed to Edwards so that customer awareness can be confirmed. International accounts were notified on 07/14/2008.
Initiated
2008-07-14
Root cause
Process control
Terminated
2008-12-23
Distribution
Worldwide Distribution --- USA including states of AL, CA, LA, MI, MN, MO, MS, NC, ND, SC, TX, and WA, and countries of Brazil, Israel, Japan, and Netherlands.
Recall number
Z-2402-2008
Recall status
Terminated
CDRH record ID
72607
FDA classified
2008-09-26
Recall event number
49049