Skip to content

Public FDA record

Z-2402-2011: Cholestech LDX Lipid Profile"GLU test cassettes, Catalog Number 10-991, Lot # 245104 Cholestech LDX Lipi

Recall record Z-2402-2011 for Biosite Inc Dba Innovacon Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2402-2011
Company in FDA record
Biosite Inc Dba Innovacon Inc.
FDA classification date
Jun 1, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated due to an issue related to an error in the encoding of calibration coefficients in the sample cassette that cause the Cholestech LDX analyzer to report glucose values as >500 mg/dL regardless of the true value of the specimen.

Source facts

Action
Alere San Diego, Inc ( formerly known as Biosite Incorporated sent an URGENT MEDICAL DEVICE RECALL letter on February 18, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. Customers were requested to: (1) Discontinue use and/or sale of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 245104. (2) Discard in accordance with your local regulations, all units of the above listed product. (3) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. (4) If you have forwarded the product listed above to another laboratory, please provide a copy of this letter to them. (5) Complete and FAX the enclosed Verification Form within 10 days to confirm your receipt of this notice, to indicate the number of boxes of test cassettes discarded from your inventory and replacement product will be sent to you. For questions about the information contained in this notification, customers were instructed to contact Alere Technical Services at 1-877-308-8289 (24 hours a day, 7 days a week) or via e-mail at service.cholestech@alere.com.
Initiated
2011-02-18
Root cause
Pending
Terminated
2012-04-02
Distribution
Worldwide Distribution - USA ( nationwide ) including the countries of South America ( EL Salvador) , Europe ( Denmark, France, Norway and Switzerland), Bahrain, Kuwait, South Africa, and United Arab Emirates.
Recall number
Z-2402-2011
Recall status
Terminated
CDRH record ID
99591
FDA classified
2011-06-01
Recall event number
58557

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗