Open original FDA record ↗Recall record
- FDA identifier
- Z-2402-2019
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Aug 28, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The injector holder could potentially fall off together with the injector and the storage box.
Source facts
- Action
- 1. Take special care when rotating the injector and storage box horizontally 2. Your Siemens Healthineers service engineer will perform an on-site check to evaluate the stability of the injector arm. 3. If there is any problem with the injector arm, your Siemens Healthineers service engineer will replace the injector arm with an improved intermediate version free of charge. 4. Please ensure all users of the affected products are aware. 5. Retain this notification in your product related records and place a copy in the medical device's instructions for use. 6. If you have distributed this product, identify your customers and notify them. If you wish to request an earlier appointment or have additional questions, feel free to contact service organization at 1-800-888-7436 .
- Initiated
- 2019-07-31
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-12-02
- Distribution
- US Nationwide Distribution in the states FL KS WI NE MS OH TX NY LA IN IL OK MI GA NJ AK CT AL MD
- Recall number
- Z-2402-2019
- Recall status
- Terminated
- CDRH record ID
- 175728
- FDA record created
- 2019-08-28
- Recall event number
- 83556