Z-2402-2021: Pipeline Flex Embolization Device

Recall record Z-2402-2021 for Micro Therapeutics Inc,. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2402-2021
Company in FDA record
Micro Therapeutics Inc,
FDA record creation date
Sep 3, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube.

Source facts

Action
On 07/13/2021, the firm sent via mail courier service an "URGENT: MEDICAL DEVICE RECALL" Notification letter to customer informing them they have received reports of serious injuries and deaths in connection with delivery pushwire fracture at the distal section of the delivery system during use. This recall affects all models of the embolization devices manufactured between April - October 2019, and April - May 2020. Customers are instructed to: 1. Do NOT use any impacted product. Remove and quarantine all unused impacted products in their inventory. 2. Return the impacted products to the Recalling Firm. Their Recalling Firm representative can assist in facilitating the return of product as necessary. If alternative product is needed, their Recalling Firm representative can assist them with identifying suitable replacement product. 3. Complete the attached Customer Confirmation Form, see Attachment 2, and fax it to 1-651-367-7075 to the attention of Neurovascular Quality or email it to rs.navfca@medtronic.com. 4. Share this communication within their organization, with other organizations where impacted devices may have been transferred, and any other associated organizations that may be impacted by this action. For additional questions and/or assistance, contact the Recalling Firm's Representative or email the Office of Medical Affairs at rs.nvoma@medtronic.com.
Initiated
2021-07-13
Root cause
Device Design
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV O.U.S.: Argentina, Brazil Canada, Chile, Colombia, Ecuador, Mexico, Peru, Austria, Belgium, Canary Island, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Kazakhstan, Latvia, Macedonia, Netherlands, Norway, Poland, Russian Federation, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, United Kingdom, Brunei Darussalam, Egypt, Israel, Jordan, Lebanon, Qatar, Saudi Arabia, Syrian Arab Republic, United Arab Emirates, South Africa, Tunisia, Reunion, China, Hong Kong, Macao, Singapore, Taiwan, Australia, New Zealand, Bangladesh, India, Pakistan, Indonesia, Malaysia, Thailand, Vietnam, Japan, and Republic of Korea.
Recall number
Z-2402-2021
Recall status
Open, Classified
CDRH record ID
188691
FDA record created
2021-09-03
Recall event number
88447

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗