Z-2403-2011: Capsule Neuron UMPC is a component in the DataCaptor" Connectivity System. Catalog Number: DC-NU-MPC Ca

Recall record Z-2403-2011 for Capsule Tech Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2403-2011
Company in FDA record
Capsule Tech Inc.
FDA classification date
Jun 1, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software: Data cached on the Neuron due to server unavailability is not being delivered to the server after server availability is restored. If a compact flash failure occurs, the file system may be corrupted and result in system hangs, lock-ups or crashes. Batteries can discharge completely in a Capsule Neuron" UMPC even though it is connected to power source

Source facts

Action
Capsule Tech initiated three corrective action notifications for the Neuron" UMPC units three failure modes as follows: a) Data cached on the Capsule Neuron" UMPC due to server unavailability is not being delivered to the server after server availability is restored: i) A customer advisory was provided to a subset of consignees, along with DataCaptor software version 6.6.5 on December 1, 2010. ii) A customer advisory was provided to the remaining consignees, along with DataCaptor software version 6.6.7 on January 27, 2011. If a compact flash failure occurs, the file system may be corrupted and result in system hangs, lock-ups or crashes: i) A customer advisory was provided to all consignees on March 23, 2011, via phone call, email and FedEx;.The correction will be provided as a software update. A formal recall notification letter will be provided to all consignees alerting them to the availability of the correction once it is released, via phone call, email and FedEx. c) Batteries can discharge completely in a Capsule Neuron" UMPC even though it is connected to power. Note: the Capsule Neuron" UMPC operates normally when connected to power. i) A customer advisory was provided to all consignees on April 23, 2010, via FedEx. For questions regarding this recall contact Capsule technical support team at (800) 260-9537 or e-mail support@capsuletech.com..
Initiated
2010-08-25
Root cause
Software design
Terminated
2012-03-20
Distribution
Worldwide Distribution - USA (nationwide) and the country of Canada
Recall number
Z-2403-2011
Recall status
Terminated
CDRH record ID
99753
FDA classified
2011-06-01
Recall event number
58636

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗