Open original FDA record ↗Recall record
- FDA identifier
- Z-2403-2020
- Company in FDA record
- Merge Healthcare, Inc.
- FDA record creation date
- Jun 19, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting in a false negative.
Source facts
- Action
- IBM Watson Health Imaging, also known as Merge Healthcare, sent an "Urgent Field Safety Notice" on May 6, 2020. The letter states that it is safe to continue to use the device as is, with the understanding that the Z-Score ranges are incorrect and should not be used to make clinical determinations. Z-Scores can be removed upon request by contacting Customer Support at the contact information below. The will provide a software correction to address the issue. A representative will be in contact with distributors to ensure all affected units have been updated. All potential users of Merge Cardio versions 10.X and 11.X with Pediatric Echo Knowledge Base with Michigan or Hybrid Z-Scores versions 3.3, 3.4, 3.5 or 5.0 should receive the notice. Customers with questions regarding the notice, please contact Customer Support at WHISupport@us.ibm.com or 877-741-5369.
- Initiated
- 2020-05-08
- Root cause
- Software Design Change
- Terminated
- 2023-06-23
- Distribution
- US Nationwide distribution including in the states of IL, TX, OR, AZ, OH, TN, CA, UT, MO, GA, NM, OK, VA, IN, CT, and VT.
- Recall number
- Z-2403-2020
- Recall status
- Terminated
- CDRH record ID
- 181430
- FDA record created
- 2020-06-19
- Recall event number
- 85633