Z-2404-2008: GE Voluson E8 Ultrasound System with software versions 6.0.0, 6.0.1,6.1.0, 6.2.0, 6.2.1, 6.2.2, 6.2,3, 7

Recall record Z-2404-2008 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2404-2008
Company in FDA record
GE Healthcare
FDA classification date
Sep 22, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Inaccurate dimension measurements in M-Mode when pan-Zoom is activated in GE Voluson E8 Ultrasound Systems. Of particular concern are potential inaccuracies of fetal cardiac measurements that may impact patient safety.

Source facts

Action
Consignees were sent a GE "Urgent Medical device Correction" letter dated April 1, 22008. The letter was addressed to Healthcare Administrator/Risk Manger; Chief of Intensive Care and Director of Clinical Engineering. The letter described the safety issue; affected product details; safety instructions stating that use of the system may be continued and measurements may be made safely in M-Mode by using HD-Zoom rather than pan-Zoom; product correction stating that a GE Healthcare Rep will contact consignees in the future to schedule software updates to resolve the issue at no charge; and contact information.
Initiated
2008-04-01
Root cause
Software design
Terminated
2011-12-11
Distribution
Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV,NY, OH, OK, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT,WA, WI, WV, WY, and PUERTO RICO, and countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KOREA, KUWAIT, LATVIA, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PARAGUAY, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, and VENEZUELA.
Recall number
Z-2404-2008
Recall status
Terminated
CDRH record ID
72691
FDA classified
2008-09-22
Recall event number
49068

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗