Open original FDA record ↗Recall record
- FDA identifier
- Z-2404-2008
- Company in FDA record
- GE Healthcare
- FDA classification date
- Sep 22, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Inaccurate dimension measurements in M-Mode when pan-Zoom is activated in GE Voluson E8 Ultrasound Systems. Of particular concern are potential inaccuracies of fetal cardiac measurements that may impact patient safety.
Source facts
- Action
- Consignees were sent a GE "Urgent Medical device Correction" letter dated April 1, 22008. The letter was addressed to Healthcare Administrator/Risk Manger; Chief of Intensive Care and Director of Clinical Engineering. The letter described the safety issue; affected product details; safety instructions stating that use of the system may be continued and measurements may be made safely in M-Mode by using HD-Zoom rather than pan-Zoom; product correction stating that a GE Healthcare Rep will contact consignees in the future to schedule software updates to resolve the issue at no charge; and contact information.
- Initiated
- 2008-04-01
- Root cause
- Software design
- Terminated
- 2011-12-11
- Distribution
- Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV,NY, OH, OK, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT,WA, WI, WV, WY, and PUERTO RICO, and countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KOREA, KUWAIT, LATVIA, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PARAGUAY, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, and VENEZUELA.
- Recall number
- Z-2404-2008
- Recall status
- Terminated
- CDRH record ID
- 72691
- FDA classified
- 2008-09-22
- Recall event number
- 49068