in case a system error occurs and the system enters the "Bypass Fluoro" mode while the X-ray locking function is active, the only way to exit the X-ray locking function, would be to either resolve the root cause of the system being in "Bypass Fluoro" or to restart the system.
Source facts
Action
A customer advisory notice, dated June 30, 2015, was sent to consignees informing them of the corrective action that would be performed to prevent the possible loss of the image generation function. Siemens plans to contact end users to install a software patch. For questions regarding this recall call 610-219-6300.
Initiated
2015-06-30
Root cause
Software design
Terminated
2016-12-05
Distribution
Nationwide Distribution
Recall number
Z-2404-2015
Recall status
Terminated
CDRH record ID
139166
FDA classified
2015-08-18
Recall event number
71762