A foot switch pedal may get stuck in the active position when the user releases the pedal,
resulting in the emission of unintended radiation
Source facts
Action
Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide actions that should be taken by the customer/user in order to prevent risks to patients. Philips will contact customers to schedule a visit to inspect the foot switch and provide an addendum to the Instructions for Use (IFU) of the system detailing the appropriate cleaning and maintenance instructions. (Philips reference 2023-IGT-BST-011).
Initiated
2023-07-19
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide Distribution
Recall number
Z-2404-2023
Recall status
Open, Classified
CDRH record ID
202718
FDA classified
2023-09-07
Recall event number
92855
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.