Open original FDA record ↗Recall record
- FDA identifier
- Z-2405-2012
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Sep 19, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
Source facts
- Action
- Elekta sent an Important Field Safety Notice dated August, 2012 to inform users that inspection and possibly bolt replacement needs to be carried out. Users should return the signed acknowledgement to Elketa, and an inspection should be carried out under FCO20002103053. For questions customers should call +44(0)1293 654200. For questions regarding this recall call 770-300-9725.
- Initiated
- 2012-08-06
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-10-08
- Distribution
- Nationwide Distribution including AL, AZ, CA, CO, CT, FL, GA, IL, IA, MD, MA, MI, MN, NE, NV, NJ, NM, NY, ND, OH, OK, OR, PA, VA, SC, TN TX, WA, WV, WI and Puerto
- Recall number
- Z-2405-2012
- Recall status
- Terminated
- CDRH record ID
- 112034
- FDA classified
- 2012-09-19
- Recall event number
- 62917