Reason stated in the source
The recall was initiated because Newport Medical has confirmed a limited number of power cords recently received from their supplier, Glob-Tek, have a slightly oversized plug connector which prevents the cord from locking into place. These cords are fully functional as supplied. However, the cord can be easily disconnected. If the cord is disconnected, the HT70 will alert the user with a power di
Source facts
- Action
- Newport Medical sent an PRODUCT NOTICE LETTER, dated September 20, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. Customers were instructed to discard the original cord and only use the replacement cord to avoid unintentional disconnection of the power cord. Each affected Customer was sent a replacement cord. Customers with any questions regarding the Product Notice were instructed to contact the Technical Support Department at (800) 451-3111.
- Initiated
- 2010-09-20
- Root cause
- Other
- Terminated
- 2011-06-03
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of MA, PA, OH, IA and Puerto Rico and the countries of Colombia, UAE, Hong Kong, Hungary, India, Japan, Nepal, China, Philippines, Singapore and South Africa.
- Recall number
- Z-2406-2011
- Recall status
- Terminated
- CDRH record ID
- 99719
- FDA classified
- 2011-06-02
- Recall event number
- 58619