Z-2406-2019: Dual-Sided Interhemispheric, Grid, Intraoperative, Strip, Wyler Subdural Electrodes

Recall record Z-2406-2019 for Ad-Tech Medical Instrument Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2406-2019
Company in FDA record
Ad-Tech Medical Instrument Corporation
FDA record creation date
Aug 28, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Supplemental information provided with devices indicates that the subdural electrodes, depth electrodes, and anchor bolts are approved for use in the MR environment, for which they are not approved for such use.

Source facts

Action
Ad-Tech Medical Instrument Corporation notified customers on about 06/18/2019, via "MEDICAL DEVICE RECALL" letter. The letter notified customers that the recall included supplemental information and not the actual physical device. Supplemental information includes: 1. Catalog - MKTG-3006 Product Catalog; 2. Attachment/Files: - Subdural MRI Certification, - Subdural MRI Testing Scheinder, - Subdural RF Heating Kanal, - Depth Electrode MRI Certification, - Depth MRI Testing Scheinder, - Depth RF Heating Kanal; 3. Journal Articles: Carmichael_NI_2010 - Feasibility of simultaneous intracranial EEG-fMRI in humans: A safety study, Carmichael_NI_2012 Simultaneous EEG-fMRI in humans: Protocol considerations and data quality, Vulliemoz_NI_2011 Simultaneous intracranial EEG and fMRI of interictal epileptic discharges in humans. Actions to be taken by the customer included to immediately examine records for any documentation subject to the recall for affected devices, and if any of the affected documentation has been further distributed, to identify those customers/personnel and immediately notify them of the recall. A copy of the recall notification letter can be included in that notification. Customers were also requested to complete the provided response form to acknowledge understanding that the devices are not approved for use as described in the recalled supplemental information. The response form also requested that customers acknowledge that the supplemental information has been properly destroyed and/or dispose of. The response form should be returned to the Ad-Tech Regulatory Team at FAX: 262-634-5668, Telephone: 262-634-1555, Email: Regulatory@adtechmedical.com. Questions or issues can also be directed to this team at the above contacts Monday - Friday 7:00 AM to 5:00 PM, Central Time.
Initiated
2019-06-18
Root cause
Employee error
Terminated
2023-01-20
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Worldwide distribution to Albania, Argentina, Australia, Austria, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Malaysia, Mexico, Netherlands, Nicaragua, Pakistan, Panama, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Syria, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam.
Recall number
Z-2406-2019
Recall status
Terminated
CDRH record ID
173935
FDA record created
2019-08-28
Recall event number
83179

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗