Z-2406-2020: MEVION S250i, MEVION S250; Proton Radiation Treatment System

Recall record Z-2406-2020 for Mevion Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2406-2020
Company in FDA record
Mevion Medical Systems, Inc.
FDA record creation date
Jun 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned off

Source facts

Action
Mevion issued letter (pCAR0012) on May 15, 2020 via mail or FedEx, with return receipt confirmation. Letter states reason for recall, health risk and action to take: Until a software fix is applied for this problem, the user is cautioned to carefully pay attention to the 3 indicators of Couch corrections. If they disagree with the intended moves the user may simply repeat the alignment step and the problem should clear. Required User Action When using multimodal imaging the user should always check the indicators for Couch Corrections before they are applied. This includes the Verity Couch Correction widget, the Verity dialog box the informs the user of the actions being sent to the Couch, and the In Room Monitor lower right panel that displays the moves before they are made. If these disagree with the intended moves, the user should back up a step and resend the corrections. Users must confirm positioning for every applied Couch correction that all intended moves are indicated and made.
Initiated
2020-05-15
Root cause
Software design
Terminated
2020-12-11
Distribution
Distribution to US in MO, NJ, OH, OK, FL, and DC, and the Netherlands
Recall number
Z-2406-2020
Recall status
Terminated
CDRH record ID
181674
FDA record created
2020-06-22
Recall event number
85748

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗