Skip to content

Public FDA record

Z-2406-2023: Constellation Procedure Pak

Recall record Z-2406-2023 for Alcon Research, LTD.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2406-2023
Company in FDA record
Alcon Research, LTD.
FDA classification date
Aug 16, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

Source facts

Action
Customers were sent a Voluntary Medical Device Correction Notification by mail and dated 06/29/2023. Customers are asked to follow the risk mitigation precautions detailed in the notice to avoid the instance of overtightening the Phaco Tip and generating plastic particles. Additionally, customers are asked to forward the recall notification to all departments or organizations that use recalled devices. Customers are asked to complete and return the provided Response Form by fax at 817-302-4337 or email at Market.Actions@alcon.com. Customers who have experienced any adverse events or quality related issues are to contact alcon via email at msus.safety@alcon.com or by phone at 1-800-757-9780, Option 5.
Initiated
2023-06-28
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, Albania, Algeria, Andorran, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Botswana, Brazil, Brunei Daruss., Bulgaria, Burma, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, French Guayana, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Isle of Man, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saint Martin, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, USA, USVI, Utd. Arab Emir., Uzbekistan, Vietnam, and Yemen.
Recall number
Z-2406-2023
Recall status
Open, Classified
CDRH record ID
202273
FDA classified
2023-08-16
Recall event number
92698

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗