Reason stated in the source

Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.

Source facts

Action
On May 21, 2026, Elekta issued a "Urgent Medical Device Notification to affected to consignees via E-Mail. Elekta asked consignees to take the following actions: 1. Please review this notification and identify if you are running the effected software version listed above. 2. If an Application Communication Time Out error message appears during a BLS plan transfer, please refer to the attached appendix. 3. Please acknowledge receipt of this notification via the Elekta Care Community or complete this form and return it to Elekta immediately upon receipt, but no later than within 30 days. 4. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. 5. Advise the appropriate personnel working with this product of the content of this letter. 6. If affected product has been transferred to a different location, please ensure that location and personnel receive a copy of this notification.
Initiated
2026-05-21
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of WI, FL, LA, NY, NY, NJ, KS, FL, IA, NM, TX and the countries of Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Lithuania, Malta, Mexico, Netherlands, Pakistan, Panama, Poland, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom.
Recall number
Z-2406-2026
Recall status
Open, Classified
CDRH record ID
220340
FDA classified
2026-06-11
Recall event number
98980

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗