Open original FDA record ↗Recall record
- FDA identifier
- Z-2409-2019
- Company in FDA record
- Elekta, Inc.
- FDA record creation date
- Aug 28, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.
Source facts
- Action
- Elekta Inc. notified customers on about 08/23/2019 via "IMPORTANT FIELD SAFETY NOTIFICATION" letter. The letter informed customers that the Monaco is displaying some incorrect shift information on the Scan and Setup Reference Report and is DICOM exporting incorrect shift information. Instructions included that all shifts should be verified before treatment, that a software fix is planned. Customer were also instructed to post the notice in a place accessible to all users, advise personnel working with the device of the issue, and to complete and return the Acknowledgement Form provided.
- Initiated
- 2019-08-23
- Root cause
- Device Design
- Terminated
- 2024-06-28
- Distribution
- Distribution within the US to Ohio and Indiana. International distribution to Germany, Greece, Sweden.
- Recall number
- Z-2409-2019
- Recall status
- Terminated
- CDRH record ID
- 175901
- FDA record created
- 2019-08-28
- Recall event number
- 83604