Z-2409-2024: NIM Vital Nerve Monitoring systems (all serial numbers manufactured) with NIM Vital System versions: v1.0

Recall record Z-2409-2024 for Medtronic Xomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2409-2024
Company in FDA record
Medtronic Xomed, Inc.
FDA classification date
Jul 24, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential for false negative response on the NIM Vital Nerve Monitoring System

Source facts

Action
Medtronic Xomed notified consignees on 06/24/2024 via letter. The letter instructed consignees that Medtronic has developed a NIM" Vital software version 1.5.4 to address this issue. Consignees were also instructed to identify affected products within inventory, share this communication within your organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action, and maintain a copy of this letter for your records. The respective Medtronic Representative will contact you to install the new software version 1.5.4 for correction of the impacted product in your possession. Additionally, consignees were asked to complete and return the customer confirmation form enclosed via email to neuro.quality@medtronic.com. Medtronic Xomed sent consignees a follow up letter with updated information on 08/22/2024. The update provided additional information and instructions to mitigate the risk with affected devices.
Initiated
2024-06-24
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic Of, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
Recall number
Z-2409-2024
Recall status
Open, Classified
CDRH record ID
208537
FDA classified
2024-07-24
Recall event number
94876

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗