Z-2410-2019: DRG Salivary Progesterone HS ELISA- IVD for the determination of progesterone in human saliva REF: SLV-59

Recall record Z-2410-2019 for DRG Instruments GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2410-2019
Company in FDA record
DRG Instruments GmbH
FDA record creation date
Aug 28, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Salivary Progesterone assay may observe a higher percentage of samples with low results. Saliva samples in the low measuring range of the kit {< 20 pg/ml) were assayed and producing a significantly lower result.

Source facts

Action
DRG International issued notification letter via email dated 5/22/19 advising of the problem, health risk and Action to be taken by Distributors and End-users: 1. We kindly ask you to immediately examine your stock and promptly quarantine these products. 2. Please send back or discard all unused kits 3. Please use the attached response form, fill, sign and send back to us 4. Please inform your customers and forward this letter and notice to persons who received the affected lot and request they fill out the form and return at once. 5. Patient samples that have been determined with the affected lot should be checked again carefully. If results are lower than expected samples should be re-run again. 6. Please tell us if you have experienced any problem with this lot and specify the details. Reply to DRG (email: qa@drg-international.com and copy drgsupport.@drginternational.com).
Initiated
2019-06-06
Root cause
Under Investigation by firm
Terminated
2020-05-11
Distribution
Distribution to Canada
Recall number
Z-2410-2019
Recall status
Terminated
CDRH record ID
173932
FDA record created
2019-08-28
Recall event number
83177

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗