Skip to content

Public FDA record

Z-2411-2011: Eleganza 3 Hospital Bed Intended use: AC Powered adjustable Hospital Bed

Recall record Z-2411-2011 for Linet Americas Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2411-2011
Company in FDA record
Linet Americas Inc
FDA classification date
Jun 3, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Linet has been made aware that there has been breakage of the Mobi-lift handle on a few beds which may cause failure during patient assist use.

Source facts

Action
The firm, Linet America, notified US customers only via phone on May 5, 2011. A follow up "service advisory information" letter dated May 11, 2011 was sent to all customers. The letter described the product, problem and action to be taken. The customers were instructed to remind/retrain staff to ensure that the Mobi-lift handles are always in the stowed position during patient transport and after use, and complete and return the attached Inspection procedure for each affected Eleganza 3 bed via mail to: Linet Americas, 10420-R Harris Oaks Blvd., Charlotte, NC 28269 or fax to: 704-248-5655. Linet Americas will be receiving a supply of new Mobi-lift design and will begin the process of executing service calls to replace the Morbi-lift handles with the new design. If you have any questions, contact the Director of Quality and Service Operations at 704-248-5650 or 980-406-6446.
Initiated
2011-05-02
Root cause
Device Design
Terminated
2012-07-19
Distribution
Worldwide distribution: USA (nationwide) including states of: KY, MA, NC, and NY; and country of: Canada.
Recall number
Z-2411-2011
Recall status
Terminated
CDRH record ID
100295
FDA classified
2011-06-03
Recall event number
58785

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗