Open original FDA record ↗Recall record
- FDA identifier
- Z-2411-2018
- Company in FDA record
- Fort Defiance Industries, LLC
- FDA record creation date
- Jul 10, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.
Source facts
- Action
- Fort Defiance Industries (FDI) notified the consignees by email on 05/24/2018. The letter explained the issue. FDI has revised the (1) Installation, Operational, and Performance Qualification (IQ/OQ/PQ) activities, (2) preparation for transport / storage, (3) preventive maintenance, and (4) device warning statements. FDI has revised the procedures for safety switch verification during IQ/OQ/PQ and complete water drainage before transport / storage of the device.
- Initiated
- 2018-05-24
- Root cause
- Device Design
- Terminated
- 2023-06-27
- Distribution
- US Nationwide Distribution in the states of PA. VA. TX
- Recall number
- Z-2411-2018
- Recall status
- Terminated
- CDRH record ID
- 165069
- FDA record created
- 2018-07-10
- Recall event number
- 80293