Open original FDA record ↗Recall record
- FDA identifier
- Z-2412-2011
- Company in FDA record
- R4 Vascular Inc
- FDA classification date
- Jun 6, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
r4 Vascular, Inc. is conducting a voluntary recall of specific lots of Duraspan Dialysis Catheters. Duraspan catheters are 15.5 Fr dual lumen chronic hemodialysis catheters. Over a six month period, seven patients at five hospitals experienced separation of the luer connector from the extension tube on the Duraspan catheter. Testing during failure analysis disclosed that these lots and other lot
Source facts
- Action
- R4 Vascular Inc sent an Urgent Medical Device Recall letter dated May 3, 2011, to all affected customers. The letter described the problem and the product being recalled. Consignees were instructed to stop distributing the product and to quarantine the recalled lots and if they had already distributed the affected product to seek its return from their customers. A template letter to physicians and healthcare facilities was enclosed for Distributors/Retailers to send or hand deliver to their customers. For questions regarding this recall call 1-704-905-2105.
- Initiated
- 2011-05-03
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-07-27
- Distribution
- Nationwide Distribution including CA, CO, CT, FL, GA, IL, LA, MD, MA, MN, MS, MO, NJ, NC, OH, PA, SC, TX, and WA.
- Recall number
- Z-2412-2011
- Recall status
- Terminated
- CDRH record ID
- 100192
- FDA classified
- 2011-06-06
- Recall event number
- 58735