Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA classification date
Jul 10, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
Source facts
Action
On 21-Jun-2018, Medtronic began notifying US physicians and health care professionals of this issue and provided Patient Management recommendations to ensure patient safety.
A Software Update to SW040 Version 8.1 will be deployed to all Medtronic 2090 and Encore clinic and hospital programmers via the Software Distribution Network or USB. Field Representatives with customers who have Percepta CRT-P patients will be sent a USB device loaded with the updated software.
In consultation with an Independent Physician Quality Panel, Medtronic recommends the following actions for physicians following patients:
- Contact your local Medtronic Representative and schedule installation of the updated Percepta CRT-P Application Software (SW040 Version 8.1) onto Medtronic 2090 and Encore clinic and hospital programmers.
- For a patient whose Percepta CRT-P device has experienced a Reset Alert or Observation: Consider scheduling an in-clinic device interrogation as soon as possible for the patients device to receive the automatic update.
- For a patient whose Percepta CRT-P device has not experienced a Reset Alert or Observation: At their next scheduled in-clinic device interrogation, the patients device will receive the automatic update.
Please complete the enclosed Clinician Confirmation Certificate and return via email to RS.CFQFCA@medtronic.com.
For questions, contact your local Medtronic Representative or Medtronic Technical Services at 800-505-4636.
Initiated
2018-06-21
Root cause
Device Design
Terminated
2020-08-19
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-2412-2018
Recall status
Terminated
CDRH record ID
165486
FDA classified
2018-07-10
Recall event number
80433