It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within
the set acceptance criteria (repeatability within 4%).
Source facts
Action
Palodex Group Oy is preparing a Class IlIl field correction to update the firmware of
ORTHOPANTOMOGRAPH OP 3D Pro of the 27 defective units and these devices
will be retested for Pan ADC reproducibility. All affected persons will be notified of
this defect in accordance with 21 CFR 806.10.