Z-2415-2024: ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting a

Recall record Z-2415-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2415-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Jul 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery

Source facts

Action
Olympus issued URGENT: MEDICAL DEVICE REMOVAL Letter on 6/21/24. Letter states reason for recall, health risk and action to take: An Olympus representative will reach out to you to coordinate a repair to replace the power supply on your affected ESG-410 unit(s). Technical analysis of this issue performed by Olympus has determined that when the device is successfully powered on, it will work within specification. Therefore, you may continue to use your ESG-410 generator until the power supply is replaced. To minimize procedural delays, power on the generator prior to starting the procedure and, whenever possible, have a replacement device available. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. If you have further distributed this product, identify your customers, and forward them this notification. 3. Acknowledge receipt of this letter through the Olympus web portal: a. Go to https://olympusamerica.com/recall Olympus requests that you report any complaints, including the ESG-410 generator not powering on, to the Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1). If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 from Monday through Friday, 9 am to 5 pm EDT, or by e-mail at Cynthia.Ow@olympus.com.
Initiated
2024-06-21
Root cause
Component design/selection
Distribution
US Nationwide distribution.
Recall number
Z-2415-2024
Recall status
Open, Classified
CDRH record ID
208574
FDA classified
2024-07-23
Recall event number
94885

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗