Open original FDA record ↗Recall record
- FDA identifier
- Z-2415-2025
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Aug 22, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
Source facts
- Action
- Customers were contacted via telephone beginning 7/14/25. The included information about the product, the issue and the action to be taken by the customers. Baxter Healthcare is requesting the return of the impacted product to perform all tests required and confirm that the product perform as expected. If you have any questions, phone (Toll Free): 800-422-9837 or 847-948-4770; hours of operation: Monday through Friday from 8:00 am to 5:00 pm CST
- Initiated
- 2025-07-14
- Root cause
- Nonconforming Material/Component
- Distribution
- US distribution to states of: AZ, FL
- Recall number
- Z-2415-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215246
- FDA classified
- 2025-08-22
- Recall event number
- 97335