Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA classification date
Jul 10, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
Source facts
Action
On 21-Jun-2018, Medtronic began notifying US follow-up physicians, and Risk Managers of health care facilities of this issue via hand delivery by Field Representatives. In consultation with an Independent Physician Quality Panel, MDT recommends the physicians following patients with affected devices:
" Consider scheduling an in-office patient follow-up as soon as possible to assess the potential for this issue per the steps described below.
" Use a programmer to ensure the Excessive Charge Time EOL&and the Low Battery Voltage ERI& Patient Alerts have been programmed to On-High
" Instruct patients to contact your office if they hear device alert tones.
" If this issue has occurred, an EOL: replace device immediately Observation will be displayed on the programmer QuickLook report. Schedule device replacement immediately.
The notification provided Patient Management recommendations to ensure patient safety and effective high voltage therapy remain as the device battery voltage approaches its 2.61V ERI threshold.
Initiated
2018-06-21
Root cause
Device Design
Terminated
2021-06-16
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-2416-2018
Recall status
Terminated
CDRH record ID
165306
FDA classified
2018-07-10
Recall event number
80379
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.