A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly
positioned cable that made contact with the housing/frame and caused a short
circuit and destroyed the fuse in the laser. This caused the shutter to remain in its
most recent state; either in an open or closed position. Once the fuse was destroyed
the shutter could not be operated.
Source facts
Initiated
2024-07-22
Root cause
Radiation Control for Health and Safety Act
Distribution
US
Recall number
Z-2416-2024
Recall status
Open, Classified
CDRH record ID
209128
FDA classified
2024-07-26
Recall event number
95028
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.