Open original FDA record ↗Recall record
- FDA identifier
- Z-2417-2012
- Company in FDA record
- Steris Corporation
- FDA classification date
- Sep 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
STERIS has learned that the Verify Bowie Dick Test Cards are not performing to product specifications. Specifically, test cards have resulted in false fail results when sterilizer performance is within acceptable ranges.
Source facts
- Action
- Steris Corporation sent a Urgent Voluntary Recall Notice dated July 3, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to examine their inventory and compare the lot numbers printed on the box with the listed lots. If they have products from the affected lot numbers to contact STERIS Customer Service for account credit. After notifying STERIS of the amount of recalled product on hand they are to immediately destroy the product. Distributors are to also notify their customers of the recall. They are to send a copy of the Customer Recall Notice. They are to inventory their stock and notify STERIS of any remaining recalled product for replacement product. All recalled product should be destroyed. Please contact your STERIS Account Manager or STERIS Customer Service at 1-800-548-4873.
- Initiated
- 2012-07-03
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-03-27
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of AL, CA, CO, CT, FL, GA, IA, IN, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK,OR, PA, RI, SC, TX, UT, VA, WA, WI, WV and WY. and the country of Canada.
- Recall number
- Z-2417-2012
- Recall status
- Terminated
- CDRH record ID
- 112673
- FDA classified
- 2012-09-20
- Recall event number
- 63129