Z-2417-2024: Arial Call Station, Model Numbers CSK200 and CSK200MR, is a wireless, battery-operated call device instal

Recall record Z-2417-2024 for Securitas Healthcare LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2417-2024
Company in FDA record
Securitas Healthcare LLC
FDA classification date
Jul 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a risk that alarms will not be received by the Arial Wireless Emergency Call System and premature depletion of the battery.

Source facts

Action
Securitas Healthcare issued a recall notice on 06/19/2024 to its consignees via USPS certified mail. The notice explained the problem, potential hazard, and requested the following: "Customer Required Actions: 1. Reduce the Missing Device Timeout to 125 Minutes Hours for all Call Stations. Clients should reduce the Missing Device Timeout to 125 minutes so that a Missing Device Alert will be triggered if a Call Station fails to send a supervision message to the Arial Emergency and Nurse Call System within 125 minutes from the last supervision message. Instructions for reducing the Missing Device Timeout to 125 minutes are attached hereto as Appendix 1. 2. Closely Monitor Missing Device Alerts within the Arial Web Application and Immediately Verify Status of Missing Devices and Cease Using any Call Stations That Do Not Reconnect. Clients should closely monitor the "Missing Device Alerts within the Arial Web Application and immediately check the status of Call Stations that trigger a Missing Device Alert and replace the Call Station battery if it is low or fully depleted. Instructions for replacing the Call Station battery are attached hereto as Appendix 2. Instructions for reconnecting missing devices and confirming that the Call Stations are properly transmitting information to the Arial Emergency and Nurse Call System are attached hereto as Appendix 3. Clients should immediately cease using any Call Station that will not reconnect to the Arial Emergency and Nurse Call System. 3. Closely Monitor the Battery Status of all Call Stations in the Arial Web Client and Replace Battery if it is Low . Clients should closely monitor the Call Station Battery Status in the Arial Web Client and immediately replace the battery if it low as described in the documentation. 4. Use Supplementary Communication Devices or Procedures for High Risk Residents. Clients should enact backup procedures, including increased room checks, or providing alternate emergency
Initiated
2024-06-20
Root cause
Software Manufacturing/Software Deployment
Distribution
US: NE, MN, WA, MD, TX, ON, WI, OR, MS, CA, and Canada
Recall number
Z-2417-2024
Recall status
Open, Classified
CDRH record ID
208563
FDA classified
2024-07-23
Recall event number
94878

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗