Z-2418-2020: Lokomat Pro, Catalog Numbers 30003 LokomatPro L6.2 120V AD ; 30004 LokomatPro L6.2 120V Kombi; 30005 Lo

Recall record Z-2418-2020 for Hocoma AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2418-2020
Company in FDA record
Hocoma AG
FDA classification date
Jun 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The motor controller may fail, which can potentially lead to an error in functionality of the Body Weights Support Rope, which means that the rope can go up or down in an uncontrolled way.

Source facts

Action
The firm began notifying affected customers on June 8, 2020. Customers were informed that the motor-controller in their Lokomat may fail, potentially leading to an unintended error in functionality of the Body Weights Support Rope which means that the rope can go up or down in an uncontrolled way. To minimize a risk in relation to a potential failure of the motor-controller, the firm advises you to do the following: - As indicated in the Instruction for Use, permanently monitor patients, which perform a training with the Lokomat. The scenario is especially critical for pediatric patients. - In any case of unusual behavior of the Body Weight Support Rope (e.g. the rope goes up unintendedly) immediately press the Emergency button on the handrail or on the Tele Stop in order to stop the treadmill and the orthosis. Immediately afterwards switch off the power supply in order to stop the rope going up. Afterwards release the patient with the emergency release knob. See Details attached. - Make sure that all possible users of the Lokomat in your institution are informed of this temporary precautionary approach. The firm will continue to work on a technical solution to further minimize the risk of the potential motor control failure.
Initiated
2020-06-08
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, GA, IL, MI, MN, MO, NE, NY, TX.
Recall number
Z-2418-2020
Recall status
Open, Classified
CDRH record ID
181951
FDA classified
2020-06-23
Recall event number
85857

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗