The reason for the field action is that the inner diameter of the Bur Guards may be out of specification. If there is excess material on the inside of the Bur Guard at the location where the guard tapers, it may come into contact with the Nose Cone of the drill. This could result in an injury or burn to the patient and/or user.
Source facts
Action
Stryker Instruments sent an Urgent Medical Device Recall Notification dated June 10, 2010, to their customers. The letter described the product the problem and the action to be taken by the customers. The customers were instructed to:
1) immediately review this recall notification.
2) forward this letter and the attached Business Reply Form (BRF) to all affected locations.
3) Immediately check and quarantine any affected product found.
4) Destroy all affected product by breaking off at least 3 of the tabs on the bur guard and then dispose of the affected product in normal trash.
5) Complete and sign the BRF and fax to Stryker Instruments Regulatory Department, 866-521-2762.
NOTE: Upon receipt of the BRF, a replacement bur guard will be sent to the customer's account for each affected bur guard destroyed.
For questions regarding this recall please contact Stryker Instruments:
1-800-800-4236, ext. 3808
jennifer.mars@stryker.com
Initiated
2010-06-02
Root cause
Device Design
Terminated
2011-04-21
Distribution
Worldwide Distribution: US AND COUNTRIES OF SCANDINAVIA, AUSTRALIA, CANADA, CHILE, CHINA, BRAZIL, FRANCE, GERMANY, GREECE, SOUTH AFRICA, ITALY, JAPAN, MEXICO, SWITZERLAND, PORTUGAL, SINGAPORE and UK.
Recall number
Z-2420-2010
Recall status
Terminated
CDRH record ID
92243
FDA classified
2010-09-13
Recall event number
55923
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.