Open original FDA record ↗Recall record
- FDA identifier
- Z-2420-2020
- Company in FDA record
- Coltene Whaledent Inc
- FDA record creation date
- Jun 23, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
Source facts
- Action
- The firm notified affected customers on June 3, 2020 by sending out Urgent Medical Device Recall letters. Customers were informed that multiple batches of the product do not meet specification. If you received one of the affected batches, please check the laser marked batch number on the forceps. If the laser marked batch number directly on the forceps matches one of the listed numbers, you should return the forceps. The recalling firm will replace your product at no cost. If you have already used the product, please inform the firm. If you should have additional questions, please contact: Tricia Cregger, Ph.D., RAC, CBA Director of Regulatory / R&D Coltene/Whaledent, Inc. 235 Ascot Parkway Cuyahoga Falls, OH 44223 USA Tricia.Cregger@coltene.com Phone: 330-916-8817 Fax: 330-916-7093
- Initiated
- 2020-06-03
- Root cause
- Process control
- Distribution
- Pending
- Recall number
- Z-2420-2020
- Recall status
- Open, Classified
- CDRH record ID
- 181797
- FDA record created
- 2020-06-23
- Recall event number
- 85794