Skip to content

Public FDA record

Z-2421-2008: Zimmer Durom cup, Metasul Durom acetabular component, uncemented, 56/50, code P, Zimmer, Winterthur, Swit

Recall record Z-2421-2008 for Zimmer Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2421-2008
Company in FDA record
Zimmer Inc.
FDA classification date
Sep 26, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Instructions for use/surgical technique instructions are inadequate.

Source facts

Action
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
Initiated
2008-07-22
Root cause
Labeling False and Misleading
Terminated
2009-10-21
Distribution
Nationwide.
Recall number
Z-2421-2008
Recall status
Terminated
CDRH record ID
72746
FDA classified
2008-09-26
Recall event number
49079

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗