Philips Medical Systems North America Co. Phillips
FDA classification date
Sep 14, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
When using DICOM print functionality, the printed images may contain data for the wrong patient.
Source facts
Action
Philips Healthcare issued an "URGENT PRODUCT CORRECTION " letter dated March 10, 2008 to their consignees. The letter described the product, problem and action to be taken by the customer.
The letter instructed the customer to:
1) Validate that image and patient information are correct on hard copy prints.
2) Reprint images that have mixed patient information, and re-check that image and patient information are correct.
3) Do not use printed images if the information is incorrect.
The firm will issue the Field Change Order (FCO) 71200025 and the field service engineers will visit each consignee and conduct a software upgrade to solves the printing problem. The consignees may contact the Philips Call Center at 1-800-772-9377, #5 #2 and reference "FCO 71200025".
Initiated
2007-12-13
Root cause
Software design
Terminated
2010-09-14
Distribution
Nationwide Distribution: including the states of AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VT, WA, and WI.
Recall number
Z-2422-2010
Recall status
Terminated
CDRH record ID
69555
FDA classified
2010-09-14
Recall event number
47477