Skip to content

Public FDA record

Z-2422-2024: MIC-KEY* Gastric-Jejunal Feeding Tube Kit, Endoscopic / Radiologic Placement, Low-Profile GJ-Tube, REF 02

Recall record Z-2422-2024 for Avanos Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2422-2024
Company in FDA record
Avanos Medical, Inc.
FDA classification date
Jul 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The IFU was missing some risk information requested by another country to be contained in the IFU pertaining to potential complications.

Source facts

Action
The recalling firm issued separate letters to distributors and customers on and dated 5/31/2024 via email and FedEx. The letter advised them the firm has updated the IFU to include additional complications associated with the use of the AVANOS MIC-KEY* Gastric-Jejunal Feeding Tubes which is supplied with the product and provides users and patients information about the intended purpose, proper use, and risks associated with the use of the device. The letter outlined the risk information pertaining to potential complications for the affected devices listed in the letter. The letter contains a QR code for the distributor or customer to access for an electronic version of the updated IFU. A Field Correction Acknowledgment form was enclosed with the distributor and customer letters for completion and return within 3 business days upon receipt of the notification. It contains a QR code to be scanned to acknowledge receipt of the notice and to complete the customer response form online. Distributors were requested to inform their customers regarding the notification and to provide them a copy of the undated Distributor Customer Letter which outlines much of the same information that is contained in the distributor and customer notification letter.
Initiated
2024-05-31
Root cause
Error in labeling
Distribution
Worldwide distribution - US Nationwide - There was also government and military distribution and the countries of Canada and Australia.
Recall number
Z-2422-2024
Recall status
Open, Classified
CDRH record ID
207123
FDA classified
2024-07-23
Recall event number
94385

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗