Defective batteries used in the Alpha Pinnacle P10 UPS External Battery Pack may overheat and cause a strong sulfuric odor to be released and a corrosive sulfuric acid solution leak to occur.
Source facts
Action
Siemens issued an Urgent Field Safety Notice in May 2010 to customers with information on what to do if the battery leaks, overheats, or vents an odor. Firm also issued a Support Bulletin to field service personnel to inspect and replace affected units.
Initiated
2010-05-05
Root cause
Nonconforming Material/Component
Terminated
2010-10-25
Distribution
Worldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Portugal, Slovenia, Spain, and Switzerland.
Recall number
Z-2423-2010
Recall status
Terminated
CDRH record ID
91737
FDA classified
2010-09-15
Recall event number
55763