Hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging material during sterilization, which resulted in degradation of the film. Over the long term it is possible that product integrity may be compromised.
Source facts
Action
On May 10, 2018, the firm notified its affected distributor via emailed memorandum of its voluntary recall. The distributor was informed of the affected lots and quantities, and asked to discontinue distribution and sales activities related the affected products.
Initiated
2018-05-10
Root cause
Package design/selection
Terminated
2019-05-10
Distribution
US Nationwide Distribution in the state of IL.
Recall number
Z-2423-2018
Recall status
Terminated
CDRH record ID
165068
FDA record created
2018-07-11
Recall event number
80291
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.