Skip to content

Public FDA record

Z-2423-2020: AIA-900 Analyzer, Part no. 022930, UDI 04560189283992

Recall record Z-2423-2020 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2423-2020
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Jun 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Substrate Monitoring System in the AIA-900 Analyzer did not detect an empty substrate bottle. The issue may lead to erroneous immunoassay test results.

Source facts

Action
On May 22, 2020, the recalling firm distributed Urgent Medical Device Correction letters to affected customers. Customers were made aware of the product issue. Immediate Actions to be taken by the Customer/User " The affected AIA-900 Analyzers are listed above by serial number. Check the serial number of your analyzer to confirm that it has the affected substrate monitoring system. o If your analyzer is not listed, please complete the form and return it to Tosoh; no further action is required. o If your serial number is listed, a Tosoh service representative will contact you to arrange for replacement of the LQC board. Additionally, please take the following actions. " Replace the Substrate bottle when the volume falls below 10 mL This is approximately inch from the bottom of the bottle or just below the bottom of the label. Be sure the supply tube extends to the bottom of bottle. Be sure to discard the substrate bottle with the remaining substrate; do not pool the remaining liquid into another bottle. " Prior to starting any run, check the volume of the substrate bottle to be sure there is enough substrate to complete the run. There are approximately 50 tests remaining in the bottle once the volume falls below the bottom edge of the label. A full bottle of substrate has approximately 500 tests. " If the substrate bottle runs dry or a DL Flag appears during a run, do not report the three (3) test results prior to the DL flag, these results may be erroneous. Replace the empty substrate bottle, repeat the Daily Check and re-run the three tests. " At the end of an assay run, check the volume of the substrate to ensure that there is fluid remaining in the bottle and the supply tube is properly submerged in the liquid. " File this notice with your laboratory records and forward this information to other supervisors or managers in your laboratory to ensure that they are aware of the potential issue. Should you have any questions regarding this m
Initiated
2020-05-22
Root cause
Component change control
Terminated
2024-08-29
Distribution
Domestic distribution nationwide. Foreign distribution to Chile, Costa Rica, Cayman Islands, Dominican Republic, Ecuador, Guatemala, Honduras, Peru, Republic of Panama, El Salvador, and Uruguay.
Recall number
Z-2423-2020
Recall status
Terminated
CDRH record ID
181680
FDA record created
2020-06-23
Recall event number
85753

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗